• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY; INTRAVASCULAR CATHETER Back to Search Results
Model Number 386807
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd cathena¿ safety iv catheter with bd multiguard¿ technology there was leakage at the catheter and hub junction.The following information was provided by the initial reporter.The customer stated: "after catheter placement, the infusion was started and leakage was then observed.When the hcp checked the product, the leakage was occurring around the area between the catheter and the catheter hub.".
 
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Additional Manufacturer Narrative
H6: investigation summary three photos and two samples, one actual and one representative without packaging, were received by our quality team for evaluation.From the photos, leakage was observed at the end of the catheter tubing near the nose of the adapter.Both samples were subjected to the catheter leak test.The actual sample failed the leak test by leaking at the end of the catheter tubing near the nose of the adapter, while the representative sample passed.The actual sample was subjected to visual inspection for tubing damage.Two cuts were observed near the nose of the adapter.A device history record could not be evaluated as the lot number is unknown.The assembly process was reviewed, there is no station that has contact with the catheter tubing at this location which will cause the observed cuts on the catheter tubing.The cuts could be caused when the product was manipulated during product application.As the sample has been used, the actual root cause could not be established.H3 other text : see h10.
 
Event or Problem Description
It was reported when using the bd cathena¿ safety iv catheter with bd multiguard¿ technology there was leakage at the catheter and hub junction.The following information was provided by the initial reporter.The customer stated: "after catheter placement, the infusion was started and leakage was then observed.When the hcp checked the product, the leakage was occurring around the area between the catheter and the catheter hub.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12976874
Report Number8041187-2021-01039
Device Sequence Number20031114
Product Code FOZ
UDI-Device Identifier00382903868070
UDI-Public00382903868070
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172506
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number386807
Device Catalogue Number386807
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2021
Supplement Date Received by Manufacturer01/10/2022
Initial Report FDA Received Date12/10/2021
Supplement Report FDA Received Date01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-