MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 17MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 17MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40017
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hyperplasia (1906); Stenosis (2263)
Date of Event 11/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is unavailable to manufacturer.
 
Event or Problem Description
It was reported that on the patient follow up, it was determined that the patient developed endoluminal intimal hyperplasia.The location of the implant was in the vasculature ica (internal carotid artery).Upon multi physician conference, it was suggested that the subject surpass evolve stent could be the cause.The patient¿s current condition is asymptomatic and there was not any mechanical injury to artery at the location of the hyperplasia via angiography.No other information was provided.
 
Additional Manufacturer Narrative
B2 adverse event/product problem - corrected : no 'product problem'.B2 outcomes attributed to ae - corrected - no 'hospitalization-initial or prolonged '.Although the dhr (device history record) could not be reviewed because the lot number for the surpass evolve remains unknown, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.It was reported on the patient follow up, the patient developed endoluminal intimal hyperplasia.Based upon medical review, the harm observed in this complaint is anticipated in nature as per the device risk assessment.As a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted within the dfu, product labeling and/or risk documentation files, an assignable cause of anticipated procedural complication will be assigned to this complaint.
 
Event or Problem Description
It was reported that on the patient follow up, it was determined that the patient developed endoluminal intimal hyperplasia.The location of the implant was in the vasculature ica (internal carotid artery).Upon multi physician conference, it was suggested that the subject surpass evolve stent could be the cause.The patient¿s current condition is asymptomatic and there was not any mechanical injury to artery at the location of the hyperplasia via angiography.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 17MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key12977690
Report Number3008881809-2021-00522
Device Sequence Number11816324
Product Code OUT
UDI-Device Identifier07613327386530
UDI-Public07613327386530
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFDS40017
Device Catalogue NumberFDS40017
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/16/2021
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date12/10/2021
Supplement Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age69 YR
Patient SexFemale
-
-