| Model Number |
96-6515 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Spring on instrument broken due to wear.New tray opened to get replacement instrument.2 minutes delay to surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that no adverse consequences were reported.Instrument broke into 2 pieces and all pieces were retrieved.The spring is required to squeeze the pin puller and so does not function without it.
|
| |
|