• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problems Environmental Compatibility Problem (2929); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the inner power line of the power cable was damaged.To restore functionality, the cable was replaced.The navigation system then passed the system checkout and was found to be fully functional.Other relevant device(s) are: product id: 9 734729, serial/lot #: unk.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system.It was reported that the power cable had a defect on contact.Sometimes it was not possible to start up the system or during a procedure there was no power and the system would run on battery power.The issue was found during preparation.Troubleshooting was performed and it was found that by bending the cable the system would sometimes power on/off.The probable cause of the reported issue was that the site was rolling-up the cable to tight causing the damage.Additional information was received.It was clarified that electricity could not be delivered from the power cable to the navigation system, resulting in the navigation system powering down and relying on the battery power.It was noted that no patient was present as the issue occurred when setting up without a patient present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12977772
MDR Text Key284828756
Report Number1723170-2021-02863
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Device Lot NumberN29925867
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-