| Model Number |
788726 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced ureteral reflux, pain or discomfort, urinary symptoms, stent dislodgement, migration, stent encrustation while using the ureteral stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is inconclusive because no sample was returned.The device was used for treatment purposes.However, it is unknown if the device had met relevant specifications or contributed to the reported event.A potential root cause for this event could be, "material selection" or "part geometry".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "potential complications associated with retrograde/antegrade positioning of indwelling ureteral stents include the following: ¿ edema ¿ stone formation ¿ peritonitis ¿ extravasation ¿ ureteral reflux ¿ stent dislodgement, fragmentation, migration, occlusion ¿ fistula formation ¿ loss of renal function ¿ hemorrhage ¿ pain/discomfort ¿ stent encrustation ¿ hydronephrosis ¿ perforation of kidney, renal pelvis, ureter and/or bladder ¿ ureteral erosion ¿ infection ¿ urinary symptoms" "with any ureteral stent, migration is a possible complication, which could require medical intervention for removal.Selection of too short a stent may result in migration." "determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort." the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced ureteral reflux, pain or discomfort, urinary symptoms, stent dislodgement, migration, stent encrustation while using the ureteral stent.
|
| |
|
Search Alerts/Recalls
|