Model Number CD2257-40
Medical Device Problem Codes Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 11/17/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to the emergency room for a follow-up after receiving inappropriate high voltage therapy.During interrogation of the implantable cardioverter defibrillator (icd), it was noted that the icd delivered inappropriate shock for atrial fibrillations with rapid ventricular response (rvr).Programming changes were performed.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD2257-40 FORTIFY ASSURA DR ICD_IDE_US
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12981654
Report Number2017865-2021-39022
Device Sequence Number11940905
Product Code NVZ
UDI-Device Identifier05414734505543
UDI-Public05414734505543
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2015
Device Model NumberCD2257-40
Device Catalogue NumberCD2257-40
Device Lot Number4312617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2021
Initial Report FDA Received Date12/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/25/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RIGHT ATRIAL LEAD; RIGHT VENTRICULAR LEAD
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexMale
-
-