• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX*FX05RE
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Health professional- requested, not provided.Occupation- requested, not provided.Pma/510(k)- k130280.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record and product-release judgement record of the involved product code/lot# combination was conducted with no findings.(b)(4).
 
Event or Problem Description
The user facility reported that the capiox device was used pre-treatment.The connector of sample line could not be fixed, easy to dis-connect during the priming.The procedure outcome was not reported.The patient was not harmed.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide a correction to h4.The incorrect device manufacture date was initially reported.Therefore, the correct date has been provided.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.2 to provide the device return date in section d9, update section h3, and to provide the completed investigation results.The actual sample was received for evaluation.Visual inspection of the entire actual sample did not find any anomaly including a breakage.Each connection part of the sampling line of actual sample was confirmed.It was found that the lock adapter on the red cock side of sampling system was run idle.Magnifying inspection of the lock adapter on the sampling system side of actual sample found that the screw thread had been deformed.Magnifying inspection of the female connector connected to the red cock side of sampling system found that the flange had been deformed.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.It was likely that since the screw thread of lock adapter and the flange of female connector were deformed, the lock adapter was run idle and could not be fixed.As a possible cause of the deformation found on the actual sample, from past experience, it was inferred that the screw thread and flange were crushed due to excessive tightening of the lock adapter.However, it was not possible to clarify when the deformation occurred from the state of actual sample.The exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX HOLLOW FIBER OXYGENATOR/ARTERIAL FILTER
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
mary o'neill
reg. no. 2243441
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key12982670
Report Number9681834-2021-00235
Device Sequence Number14866890
Product Code DTZ
UDI-Device Identifier04987350781758
UDI-Public04987350781758
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K071572
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Catalogue NumberCX*FX05RE
Device Lot Number200904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2021
Supplement Date Received by Manufacturer12/14/2021
12/13/2021
Initial Report FDA Received Date12/13/2021
Supplement Report FDA Received Date12/14/2021
01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-