• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC250
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
Provider alleges, consumer alleges, the cord allegedly caught on fire.
 
Additional Manufacturer Narrative
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
 
Event or Problem Description
Provider alleges, consumer alleges, the cord allegedly caught on fire.
 
Additional Manufacturer Narrative
The hand control harness was damaged 12" and 18" from the connector end.There was evidence of exposed wiring but no evidence of fire or shorting.The chair was still functional despite this damage.The hand control harness was damage due to customer abuse.The owner's manual contains instruction to periodically check the hand control harnessing for damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
8008008586
MDR Report Key12983291
Report Number2530130-2021-00133
Device Sequence Number17566374
Product Code INO
UDI-Device Identifier00606509400237
UDI-Public00606509400237
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K070950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberLC250
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date12/13/2021
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-