| Model Number |
60-03-75 |
| Medical Device Problem Code |
Decreased Pump Speed (1500)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump console.The incident occurred in austin, texas.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that the flow could not be increased on a centrifugal pump console during a procedure.There was no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H.10: no service activity has been scheduled yet for this event.The service is pending approval from customer.Based on investigation results on similar complaints, possible root causes of the reported event can be attributable to: - defective flow sensor board, - defective flow sensor, - defective shaft angle encoder, - loose connection on electronic boards.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|