CORDIS CORPORATION POWERFLEXPRO 6MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400604S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2021
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the balloon of a powerflex pro 6mm 4cm 80 ruptured.There was no reported injury to the patient.The device will not be returned for evaluation as it was discarded.The lesion to be treated was in the iliac artery.
 
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with lot 82220667 presented no issues during the manufacturing process that can be related to the reported event.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the balloon of a powerflex pro 6mm 4cm 80 ruptured.There was no reported injury to the patient.The device will not be returned for evaluation as it was discarded.The lesion to be treated was in the iliac artery.Additional information was requested; however, the information could not be obtained.
 
Additional Manufacturer Narrative
As reported, the balloon of a powerflex pro 6mm 4cm 80 percutaneous transluminal angioplasty (pta) balloon catheter ruptured.There was no reported injury to the patient.The lesion to be treated was in the iliac artery.Additional information was requested; however, the information could not be obtained.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82220667 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.(b)(4).Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 6MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key12984470
Report Number9616099-2021-05176
Device Sequence Number12176443
Product Code LIT
UDI-Device Identifier20705032059517
UDI-Public(01)20705032059517(17)240331(10)82220667
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number4400604S
Device Catalogue Number4400604S
Device Lot Number82220667
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2021
Supplement Date Received by Manufacturer12/15/2021
Initial Report FDA Received Date12/13/2021
Supplement Report FDA Received Date01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-