| Model Number |
24696 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.0mm x 80mm x 150cm, mr coyote balloon catheter was advanced for dilatation.However, during the procedure, the device broke where the balloon met the shaft.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis could not confirm the reported separation.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 2.0mm x 80mm x 150cm, mr coyote balloon catheter was advanced for dilatation.However, during the procedure, the device broke where the balloon met the shaft.The procedure was completed with a different device.No patient complications were reported.
|
| |
|