| Model Number |
7130 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There were numerous hypotube kinks.There was contrast in the inflation lumen and balloon.There was blood in the guidewire lumen and the balloon was loosely folded.The device was soaked for a period of time to break up the contrast and blood in the device.The device was then prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole 6mm from the tip of the device.The device failed to inflate to rated burst pressure.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 07dec2021.It was reported that balloon leak occurred.A 2.50mm x 8mm emerge balloon catheter was advanced for dilatation.However, during first inflation the device failed to inflate.When the device was removed and inflated, a leaked was notice.The procedure was completed with another of same device.No patient complications were reported.However, device analysis revealed a pinhole.
|
| |
|