| Model Number |
EUP3012X |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use a euphora rx ptca balloon catheter.There was no damage noted to the packaging.There was no difficulty noted when removing the device from the hoop.The device was inspected with issues noted.Negative prep was not performed.It was reported that the stylet was stuck in the tip of the device.The stylet at the end of balloon had resistance and was unable to be removed.Force was applied causing the staff to tear the glove.There was no harm caused to the staff.The device was not attempted to be loaded onto the guidewire.The device was not used in the patient.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the device returned with the stylette partially loaded in the catheter.Approximately 1.4 cm of the stylette was exposed on device return.Necking was evident to the proximal balloon bond.Necking was evident to the distal tip.The distal tip was cut radially, proximal to the necking and it was then possible to remove the stylette.Residue was visible on the section of the stylette that was inserted into the catheter.The inner lumen could not be verified due to the condition of the retuned device.No other deformation was evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|