• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P ALTIVATE REVERSE SHOULDER; ALTIVATE REVERSE, HUMERAL STEM, SMALL SHELL, SZ 6X48MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P ALTIVATE REVERSE SHOULDER; ALTIVATE REVERSE, HUMERAL STEM, SMALL SHELL, SZ 6X48MM Back to Search Results
Model Number 533-06-048
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/16/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the positioning of the humeral implant.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as the positioning of the humeral implant.The previous surgery and the surgery detailed in this event occurred 1 year and 3 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr #52673 associated with the main part #533-06-048, altivate reverse, humeral stem, small shell, sz 6x 48mm which documents that out of 15 parts lot, 1 part was rejected due to laser mark was not properly aligned.Later the rejected part was accepted to use as is after proper justification.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to position of the humeral implant.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE REVERSE SHOULDER
Common Device Name
ALTIVATE REVERSE, HUMERAL STEM, SMALL SHELL, SZ 6X48MM
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key13002035
Report Number1644408-2021-01443
Device Sequence Number14673257
Product Code KWS
UDI-Device Identifier00190446334958
UDI-Public(01)00190446334958
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190290
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/11/2025
Device Model Number533-06-048
Device Catalogue Number533-06-048
Device Lot Number1490A1033
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2021
Initial Report FDA Received Date12/14/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
509-02-032 LOT 951W1434
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-