Model Number 0998-00-0800-55
Device Problems Device Displays Incorrect Message (2591); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Manufacturer Narrative
Type of investigation not yet determined: additional information is being requested in regards to whether or not the customer has requested for getinge to service the iabp unit involved.A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that during a routine check performed by the customer the cardiosave intra-aortic balloon pump (iabp) had an audible alarm with an internal communication error.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) evaluated the iabp unit and detected fault# 139.The fse reported that the power management board would need to be replaced.The fse returned at a later date and replaced the faulty power management board which corrected the issue.The fse then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The first name of the initial reporter in block e1 has been abbreviated due to field character limit; the full name should read (b)(6).
 
Event Description
N/a.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13002485
MDR Text Key289500268
Report Number2249723-2021-02866
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-