It was reported that during acl procedure, internal to patient, the screw failed to be placed, it did not fit and broke off, the broken pieces were removed using kelly forceps.The procedure was successfully completed without significant delay using a s+n back-up device.No further complications were reported.
|
|
H10, h3, h6.The reported device was received for evaluation.It was determined the device did not contribute to the reported event.A complaint history review concluded this was an isolated event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A review of the customer provided image found two screws with damaged and flattened threads.Both screws have debris in the threads.A visual inspection of the returned device found that it is not in it's original packaging.The threads of the screw are damaged and flattened.The threads are torn and broken.There is debris on the device.Based on the condition of the product material found during visual inspection, additional material testing is not required.A functional evaluation could not be performed due to the condition in which the device was received.No clinically relevant supporting documentation was provided for review.If any fragments of the biosure screw remains retained in the patient, it is made of a biocomposite material which is intended for implantation to allow for bone ingrowth.If the screw was retained, it is unknown if the biosure screw will migrate and there is potential risk for tissue inflammation and/or pain.Since no patient injury or other complications were reported, no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|