| Model Number |
A35HPV06100080 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Image review: the customer returned one image.The image appears to show inflated balloon in a patient¿s vasculature.The middle of the balloon is not inflated which could indicate a twist in the balloon.However, this is a hypothesis and cannot be positively determined.The 2 balloon marker bands are visible in the image also.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician attempted to use an fortrex hp pta balloon catheter during patient treatment.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.No issues were noted when removing the device from the hoop/tray.Ifu was followed and the device was prepped without issue.An inflation device was used for balloon inflation.The device did not pass through a previously deployed stent.No resistance was noted during advancement.It is reported balloon inflation difficulties occurred when device was inflated to nominal burst pressure.There was no leak noted in the device.The balloon did not burst.The device was safely removed from the patient.A new fortrex device was used to complete the procedure.No patient injury reported.
|
| |
|