Model Number 10104-006
Medical Device Problem Code Environmental Particulates (2930)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported an event of a ¿mist¿ or haze emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Additional Manufacturer Narrative
Initial reporter zip code: (b)(6).Initial reporter phone number: (b)(6).A field service engineer was dispatched to the customer site and identified that the oil mist filter screw was loose.The oil mist filter was tightened/reseated to resolve the mist/haze issue.Unit meets specifications and was returned to service.Asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the mist/haze issue, and system risk analysis (sra).The device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the mist/haze issue for the sterrad® 100nx unit was reviewed for the prior six months from open date and no significant trend was observed.Review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were replaced in resolving this issue; therefore, no parts were available for further evaluation.The assignable cause of the mist/haze is likely due to the oil mist filter.The asp field service engineer tightened/reseated the part and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® 100NX ALLCLEAR STERILIZER 2-DR
Common Device Name
STERRAD® EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
gabriela mclellan
33 technology drive
irvine, CA 92618
9495030264
MDR Report Key13003966
Report Number2084725-2021-00440
Device Sequence Number11969444
Product Code MLR
UDI-Device Identifier10705037048687
UDI-Public10705037048687
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K071385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10104-006
Device Catalogue Number10104-006
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/18/2021
Initial Report FDA Received Date12/14/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-