| Model Number |
ICONAAN |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint icon auto cpap is currently is en-route to fisher and paykel healthcare(f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A distributor in (b)(4), reported that the power cord of an icon auto cpap humidifier was broken with exposed wires.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint icon auto cpap humidifier was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the customer and our knowledge of the product.Results: the customer reported that the power cord was broken with exposed wires.Conclusion: without the complaint device, we are unable to confirm the reported event.During initial assembly of the icon auto cpap, all power cords are visually inspected for damage.Once the assembly process is completed, all icon auto cpap devices are visually inspected again before release for distribution.This suggests the damage occured after it had been distributed.The icon auto cpap is designed to the electrical safety standards, ul60601-1 and as/nzs 3200.1.The materials used in the thermoplastic components of the mains connector and the cases are flame retardant according to ul 60601-1 and as/nzs3200.1.Our user instructions that accompany the icon auto cpap humidifier state the following: - "only operate if the device, power cord and plug are dry and in good working order." - "do not operate the device, water chamber or breathing tube if it is dropped, damaged or not working as intended.".
|
| |
|
Event or Problem Description
|
|
A distributor in texas, reported that the power cord of an icon auto cpap humidifier was broken with exposed wires.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|