| Model Number |
3100A |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Pneumothorax (2012)
|
| Date of Event |
10/13/2021
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported to vyaire medical that the o2 (oxygen) blender was not working properly.Ino (inhaled nitric oxide) was administered via ino delivery device with high flow oxygen blender.The blender read 100% however the analyzer read 21%.They attempted to increase oxygenation, the infant sustained a pneumothorax and health declined rapidly and the patient passed with parents.They discovered that the nut retaining the knob of the o2 blender was loose and was not providing friction fit required for adjustment of o2 concentration.The customer confirmed that the health care provider attempted to increase oxygenation.However, the infant's health sustained pneumothorax.Infant's health declining rapidly and passed with parents.
|
| |
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
|
| |
|