| Model Number |
6533 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first day of the month of the bsc aware date since event date was not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 20mm x 3.25mm nc quantum apex balloon catheter was advanced for dilatation.Balloon was intact and fine when it was opened from the package.However, upon inflation, it was fractured in the middle of the wire and completely broke off from the system.The balloon was able to be retrieved from the body with a guide extension catheter.The procedure was completed with a different device.No further patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.Date of event: used the first day of the month of the bsc aware date since event date was not provided.Device evaluated by manufacturer: returned product consisted of a nc quantum apex balloon catheter.The device was visually and microscopically examined.There were numerous hypotube kinks to the device.There was a separation of the hypotube 82.6cm from the strain relief.The separated ends were ovaled in shape indicating the device was kinked prior to separation.The rapid port exchange was torn a total length of 2mm which indicates an interaction or difficulty with a guidewire.There was contrast in the inflation lumen, blood in the guidewire lumen, and the balloon was tightly folded.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 20mm x 3.25mm nc quantum apex balloon catheter was advanced for dilatation.Balloon was intact and fine when it was opened from the package.However, upon inflation, it was fractured in the middle of the wire and completely broke off from the system.The balloon was able to be retrieved from the body with a guide extension catheter.The procedure was completed with a different device.No further patient complications nor injuries reported.
|
| |
|