| Model Number |
TN FC500 MPL FLOW CYTOMETER 100-240V 50/60HZ |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Section a patient information is not applicable.There was no impact to patients as a result of this event.The field service engineer (fse) was on site and confirmed what the customer reported.To resolve the issue the fse replaced the tarpon amp boards at the affected locations (fl1 and fl4).Bec is filing an mdr for this event based on the fda classification of the (b)(6) 2018 urgent medical device recall as a class i recall (b)(6) 2018 (recall number z-0471-2019 for fc 500; recall number z-0472-2019 for epics xl/xlmcl).Bec internal identifier (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported amp board error at the fl1 and fl4 positions on their fc 500 flow cytometer.There was no report of death, injury, or change to patient treatment as a result of this event.
|
| |
|
Search Alerts/Recalls
|