| Model Number |
TN DXH 900 HEMATOLOGY SYSTEM WITH FLOOR STAND |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was at the customer site and inspected the instrument.To resolve the issue the fse primed sweep flow and zapped apertures to resolve the issue.The information provided reasonably suggest an instrument malfunction related to dirty apertures.Bec internal identifier - (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported getting dashes on red blood cell (rbc), rbc parameters and platelet (plt) on their dxh 900 hematology instrument.The instrument was also generating 'check for clots' errors for plt.The customer also reported erroneously high plt results for two patients.Patient data was requested but the customer was unable to provide it because they did had not separated them and did not have time to seek them out.The erroneous results were not reported out of the laboratory.There was no report of death, injury, or change to patient treatment as a result of this event.
|
| |
|