| Model Number |
393-800 |
| Medical Device Problem Codes |
High Readings (2459); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the complaint on (b)(6) 2021 the customer noticed false high measurements of sodium on five patients on a abl800 flex analyzer.The customer noticed that the refence membrane was full of blood and chamber was flooded with blood and leaked into ph electrode chamber.She cleaned up the blood and replaced the membrane, ran calibrations and qc, these passed and the patients after that were good.The qc and calibrations passed in the morning, showing no indication of a problem on the analyzer.The below are the false high measurements of na+.Comparison measurement/expected value is 140mmol/l.Patient 1: 159 mmol/l.Patient 2: 170 mmol/l.Patient 3: 161mmol/l.Patient 4: 157mmol/l.Patient 5: 159mmol/l.No reports of death or serious injury.
|
| |
|
Event or Problem Description
|
|
According to the complaint on (b)(6) 2021 the customer noticed false high measurements of sodium on five patients on a abl800 flex analyzer.The customer noticed that the refence membrane was full of blood and chamber was flooded with blood and leaked into ph electrode chamber.She cleaned up the blood and replaced the membrane, ran calibrations and qc, these passed and the patients after that were good.The qc and calibrations passed in the morning, showing no indication of a problem on the analyzer.The below are the false high measurements of na+.Comparison measurement/expected value is 140mmol/l.Patient 1: 159 mmol/l.Patient 2: 170 mmol/l.Patient 3: 161mmol/l.Patient 4: 157mmol/l.Patient 5: 159mmol/l.No reports of death or serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The radiometer investigation has shown that it is plausible that the leakage on the reference membrane is the root cause for the deviating measurement performance on sodium.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The radiometer investigation is finalized and the root cause is concluded to be "supplied materials outside specification".
|
| |
|