| Model Number |
8888571562 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer states the cdu was leaking at the collection chamber.The target vacuum wasn't reached.This happened during use on a patient and there was no injury.The unit was switched with another without any issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.Three (3) unopened representative samples were received.Upon visual inspection no issues were observed.All 3 returned samples were tested as per product specifications for sentinel seal / sentinel seal dual c.D.U function test.A series of tests were performed on the 3 units and all 3 devices met all functional requirements.Based on the 3 returned samples this compliant cannot be confirmed as this time.A risk assessment was deemed not required as the investigation did not confirm that the reported issue was related to a manufacturing process.Further monitoring will be performed via trending within the scope of post market surveillance.At this time, a corrective/preventative action (capa) is not required as the complaint samples and dhr did not indicate a device related issue.
|
| |
|
Search Alerts/Recalls
|