| Model Number |
PS10.0-50D235 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Fistula (1862); Pain (1994)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that the patient developed a bony fistula and complained of pain.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon return, visual inspection of the nail revealed the distraction rod of the nail had visible evidence of discoloration at the male/female junction, which confirmed discoloration failure mode.The discoloration failure is being investigated under a capa.The nail listed in this complaint was built prior to the initiation of the capa.The device history records were reviewed, and found the devices were manufactured in accordance with the specified requirements and met all the required quality inspections.There were no discrepancies or non-conformances related to the reported event.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|