NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-50D235
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fistula (1862); Pain (1994)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that the patient developed a bony fistula and complained of pain.
 
Manufacturer Narrative
Upon return, visual inspection of the nail revealed the distraction rod of the nail had visible evidence of discoloration at the male/female junction, which confirmed discoloration failure mode.The discoloration failure is being investigated under a capa.The nail listed in this complaint was built prior to the initiation of the capa.The device history records were reviewed, and found the devices were manufactured in accordance with the specified requirements and met all the required quality inspections.There were no discrepancies or non-conformances related to the reported event.
 
Event Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key13010839
MDR Text Key282727904
Report Number3006179046-2021-00514
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00887517950741
UDI-Public887517950741
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K180503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPS10.0-50D235
Device Lot Number9052005AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/08/2021
Initial Date FDA Received12/15/2021
Supplement Dates Manufacturer Received02/11/2022
Supplement Dates FDA Received02/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age21 YR
Patient SexFemale
Patient Weight63 KG
-
-