SMITH & NEPHEW, INC. TRIGEN LOW PROFILE SCREW 5.0MM X 42.5MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
71645042 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369); Unequal Limb Length (4534)
|
| Date of Event |
07/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, after an open reduction with osteosynthesis had been performed on (b)(6) 2018 to treat an intertrochanteric femoral neck multifracture, the patient experienced pseudoarthrosis with a nonunion of the treated femoral fracture, a right leg length discrepancy, anteflexion deformity and the fracture of the intertan 11.5mm x 38cm 125d rt along with the breakage of a proximal screw.A revision surgery was performed on (b)(6) 2021 to treat this adverse event.During this procedure, the nail and screws were removed, and a hip joint replacement construct was implanted.The patient¿s outcome is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional info in: b, d, g, h results of investigation: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was fractured into two pieces.Both of the pieces were returned.A review of complaint history did not reveal similar events for the listed device.The clinical/medical evaluation concluded that the x-rays and revision operative note do confirm the reported fracture of the intertan nail, pseudarthrosis, right leg length discrepancy, anteflexion deformity, and breakage of the trigen screw.However, based on the limited information provided, the root cause of the reported events cannot be confirmed.It cannot be concluded that the reported clinical findings are associated with a malperformance of the implant but could be related to the reported non-union.The patient impact beyond the revision cannot be determined.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.The manufacturing process of the device did not contribute to the reported event.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, after an open reduction with osteosynthesis had been performed on (b)(6) 2018 to treat an intertrochanteric femoral neck multifracture, the patient experienced pseudoarthrosis with a nonunion of the treated femoral fracture, a right leg length discrepancy, anteflexion deformity and the fracture of the intertan 11.5mm x 38cm 125d rt along with the breakage of a trigen low profile screw 5.0mm x 42.5mm.A revision surgery was performed on(b)(6) 2021 to treat this adverse event.During this procedure, the nail and screws were removed, and a hip joint replacement construct was implanted.The patient¿s outcome is unknown.
|
| |
|