| Model Number |
110-120 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a4, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified one extended opening.A weight test of the device was verified and the device underweight.A microscopic analysis was performed which identified one sharp edge opening in the shell with nick and nick striated.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is a sharp edge opening in the shell with nick in the side posterior assessed as surgical impact opening, and striated nicks assessed as surgical tool scratch like.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted and replaced.
|
| |
|