| Model Number |
550AVHCT |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed.Per the srt, during oven testing when the bpm was held in a 20 degrees celsius environment it was reading 14.9 degrees celsius which is outside of the specification range.The srt replaced the venous bpm probe.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
The service repair technician (srt) reported that during testing of the device at the service center, the venous blood parameter monitor (bpm) probe failed measured temperature accuracy testing.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|