The monitor and electrode belt have not been returned for evaluation.Based on the data available at this time, there is no indication of a device malfunction causing or contributing to the inappropriate treatment.Inappropriate defibrillations are an anticipated risk associated with the use of the hospital wearable defibrillator.Patients are instructed through alarms, voice messages, ifu, and training to press the response buttons to prevent an inappropriate defibrillation.A summary of the safety and effectiveness data (ssed), including the inappropriate defibrillation safety objective supporting fda's approval of the lifevest and hwd, can be found at (http://www.Accessdata.Fda.Gov/cdrh_docs/pdf/p010030b.Pdf).The zoll detection algorithm complies with iec 60601-2-4 performance requirements for sensitivity and specificity.
|