Model Number 0998-00-3023-67
Device Problems Gas Output Problem (1266); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2021
Event Type  malfunction  
Manufacturer Narrative
At this time, the customer has not requested for getinge to evaluate the cs300 intra-aortic balloon pump (iabp) unit involved in this reported event.The getinge field service engineer (fse) advised that the customer was contacted and performed the repair to the unit involved in this reported event due to the problem being connected to the helium tank / gasket.Based on the information available at this time, the reported issue was resolved by replacing the helium tank.The iabp was then returned to clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported by the customer that during use the cardiosave intra- aortic balloon pump (iabp) unit began to sound a deafening whistle that was found to be coming from the helium cylinder.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
Additional information: e1(event site postal code: (b)(6)).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, UNKNOWN
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13013236
MDR Text Key284019974
Report Number2249723-2021-02880
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-67
Device Catalogue Number0998-00-3023-67
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-