| Model Number |
83786 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Premature Activation (1484); Stretched (1601); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device eval by mfr: the device was received for analysis.The returned device matches with upn provided by the customer.Only a main coil was returned for this complaint.The delivery wire and introducer sheath did not return.The main coil was found kinked and stretched.No more damages were found.Microscopic inspection revealed the zap tip had a smooth surface.The interlocking arm was detached.Dimensional inspection was performed and the device met specification.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 07dec2021.It was reported that the coil prematurely deployed.A.035 interlock 2d 15mm x 40cm embolic coil was selected to embolize a splenic artery aneurysm.The coil was advanced into the middle of the catheter.There was friction advancing the coil.However, the coil prematurely deployed and became stretched.The coil and catheter were removed.A new coil and catheter were used to complete the procedure.There were no patient complications.However, device analysis revealed a detached interlocking arm.
|
| |
|