| Model Number |
HLO51035 |
| Medical Device Problem Codes |
Unraveled Material (1664); Device-Device Incompatibility (2919); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 08/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the right superficial femoral artery, the sheath was allegedly hard to be inserted over the guidewire.It was further reported that the tip of the sheath was found to be allegedly damaged and peeled.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 08/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the right superficial femoral artery, the sheath was allegedly hard to be inserted over the guidewire.It was further reported that the tip of the sheath was found to be allegedly damaged and peeled.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|