For the investigation the logfile was analyzed.It was found that at the day of event the device was put into operation after a successful power on self-test.After about 45 minutes a power supply inop-alarm was generated as and the device switched to battery operation as specified.After 3 minutes on battery operation the device generated an alarm that as only 10 % battery-charge was remaining.Just one minute later, the device performed a warm start due to the power disruption, accompanied by a "ventilator failure¿ alarm.The device switched to man/spont mode as specified.In case the battery is fully charged, operation can be continued with the current settings for at least 30 minutes.Obviously, the charging level of the battery decreased faster than expected.As the batteries are subject to an aging process it can be assumed that the identified symptom most likely was caused by an used-up battery.To prevent this, the battery must be replaced every 3 years during product maintenance.Unfortunately, it is not known when the battery was replaced last time so it is not possible to make a statement about the age of the battery.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|