• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF CUP POSITIONER/IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF CUP POSITIONER/IMPACTOR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71362299
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal reference number: case-(b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, the top of a handle of a ref cup positioner/impactor broke while impacting.As reported, the device broke outside the patient and no pieces fell into the patient.The procedure was completed, without delay, using a s+n back-up device.Patient was not harmed.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device was conducted and confirmed the top of the device is broken off, rendering it inoperative.The other piece was returned with the product.A review of complaint history revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF CUP POSITIONER/IMPACTOR
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key13018540
Report Number1020279-2021-08702
Device Sequence Number4625908
Product Code JDG
UDI-Device Identifier03596010443236
UDI-Public03596010443236
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2015
Device Model Number71362299
Device Catalogue Number71362299
Device Lot Number04MM02563
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2021
Supplement Date Received by Manufacturer02/09/2022
Initial Report FDA Received Date12/16/2021
Supplement Report FDA Received Date02/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-