• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4.5 HEALIX TI 3SUTURE ANC W/OC; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4.5 HEALIX TI 3SUTURE ANC W/OC; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 222271
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Udi: (b)(4).The lot number was unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the healthcare professional through clinical research report on clinical evidence on safety and performance of the 4.5 healix ti 3suture anchor w/oc device that postoperatively to a rotator cuff repair surgery of the left shoulder performed on (b)(6) 2019, it was observed that the anchor pulled out required a revision surgery performed on (b)(6) 2019 for the removal of the loose anchors.The current status of the patient was unknown.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.This complaint file was opened to document complaints derived through a journal article review.The journal article review indicated depuy mitek's product failure(s).Multiple attempts were done to obtain more information from the author, however, no response was received.It is unknown if complaints derived from this journal article were previously reported and documented in the depuy mitek complaint system at the time of occurrence as no product code/lot number information was provided to perform the search.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported problem.Since no lot number was provided, a manufacturing record evaluation or sterile load review could not be conducted.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent to which this complaint is observed in the field.Device history lot: no lot numbers were supplied which precludes conducting a dhr review or a lot-specific search in the complaints handling system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 HEALIX TI 3SUTURE ANC W/OC
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key13018619
Report Number1221934-2021-03772
Device Sequence Number17384347
Product Code HWC
UDI-Device Identifier10886705008162
UDI-Public10886705008162
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K082282
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number222271
Device Catalogue Number222271
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2021
Supplement Date Received by Manufacturer12/22/2021
Initial Report FDA Received Date12/16/2021
Supplement Report FDA Received Date12/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-