• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S FOLYSIL CATHETER; CATHETER, RETENTION TYPE, BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S FOLYSIL CATHETER; CATHETER, RETENTION TYPE, BALLOON Back to Search Results
Model Number AA61101002
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
See manufacturer report numbers 9610711-2021-00118 and 9610711-2021-00119 regarding the other two occurrences of balloon burst.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional information prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event or Problem Description
According to the available information, three balloons burst in the pediatric patient¿s bladder when the balloon was blocked with 6 ml.The mother blocks the balloon and had used the product daily for several months.She stated the balloon can be blocked with maximum 15 ml, but it was noted she had not read the instructions for use.The patient was not harmed.
 
Additional Manufacturer Narrative
See manufacturer report numbers 9610711-2021-00118 and 9610711-2021-00119 regarding the other two occurrences of balloon burst.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional information prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event or Problem Description
According to the available information, three balloons burst in the pediatric patient¿s bladder when the balloon was blocked with 6 ml.The mother blocks the balloon and had used the product daily for several months.She stated the balloon can be blocked with maximum 15 ml, but it was noted she had not read the instructions for use.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL CATHETER
Common Device Name
CATHETER, RETENTION TYPE, BALLOON
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
Manufacturer (Section G)
COLOPLAST A/S MANUFACTURING
9 avenue edmond rostand
sarlat-la-caneda 24206
FR   24206
Manufacturer Contact
usaby angela kilian
1601 west river road north
minneapolis, MN 55411
8007880293
MDR Report Key13019001
Report Number9610711-2021-00120
Device Sequence Number5127413
Product Code EZL
UDI-Device Identifier03600040127100
UDI-Public03600040127100
Combination Product (Y/N)N
PMA/510(K) Number
K013174
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/04/2026
Device Model NumberAA61101002
Device Catalogue NumberAA6110
Device Lot Number8133788
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2021
Initial Report FDA Received Date12/16/2021
Date Device Manufactured08/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age10 YR
Patient SexMale
-
-