| Model Number |
26665 |
| Medical Device Problem Codes |
Positioning Failure (1158); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the stent was bent.A carotid wallstent was selected for use in a coronary artery stenting procedure.During the procedure, after the stent passed the lesion, the stent was not able to deploy.The stent tip was found to be bent and was then removed.The procedure was successfully completed with a different device.There were no patient complications and the patient is fine.
|
| |
|
Event or Problem Description
|
|
It was reported the stent was bent.A carotid wallstent was selected for use in a coronary artery stenting procedure.During the procedure, after the stent passed the lesion, the stent was not able deploy.The stent tip was found to be bent and was removed.The procedure was successfully completed with a different device.There were no patient complications and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: a carotid device was returned for analysis.A visual and tactile examination identified that the tip of the device was bent.The device was returned with the stent fully mounted in the correct location on the device.The investigator deployed the stent without any issue.Damage was noted on the distal end of the stent.
|
| |
|