Model Number EPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2021
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported their epiq cvxi ultrasound system shut down in echo navigator mode during a mitra clip heart valve procedure.The procedure in progress was completed successfully without impacting patient outcome.The patient was not harmed as a result of the issue.
 
Additional Manufacturer Narrative
Evaluation of the system logs and workflow will be included in a follow up report upon investigation completion.
 
Additional Manufacturer Narrative
A thorough investigation was performed to identify the root cause of the reported issue.However, the engineering team was unable to reproduce the failure.The field service engineer upgraded the system software to resolve the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ CVXI
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key13019319
Report Number3019216-2021-10159
Device Sequence Number4704370
Product Code IYN
UDI-Device Identifier00884838097940
UDI-Public00884838097940
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Device Catalogue Number795232
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/03/2021
Supplement Date Received by Manufacturer12/03/2021
Initial Report FDA Received Date12/16/2021
Supplement Report FDA Received Date11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-