| Model Number |
EPIQ DIAGNOSTIC ULTRASOUND SYSTEM |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported their epiq cvxi ultrasound system shut down in echo navigator mode during a mitra clip heart valve procedure.The procedure in progress was completed successfully without impacting patient outcome.The patient was not harmed as a result of the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the system logs and workflow will be included in a follow up report upon investigation completion.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough investigation was performed to identify the root cause of the reported issue.However, the engineering team was unable to reproduce the failure.The field service engineer upgraded the system software to resolve the reported issue.
|
| |
|