MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number W4TR01 |
Device Problem
Pacing Problem (1439)
|
Patient Problems
Dyspnea (1816); Dizziness (2194)
|
Event Date 12/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: 479888, lead, implanted: (b)(6) 2020, 5076-52, lead, implanted: (b)(6) 2020.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient was experiencing dizziness, "heavy legs", and shortness of breath.It was noted that the patient's pacing rate was low and the presenting rhythm showed bigeminy premature ventricular contractions (pvc).It was also reported that there was a possible high left ventricular (lv) lead threshold.The crt-d and lv lead remain in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient was experiencing dizziness, "heavy legs", and shortness of breath.It was noted that the cardiac re synchronization therapy pacemaker (crt-p) pacing rate was low and the patient's presenting rhythm showed bigeminy premature ventricular contractions (pvc).It was also reported that there was a possible high left ventricular (lv) lead threshold.The crt-p and lv l ead remain in use.No further patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|