Material Rupture (1546); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a severely calcified lesion in the left superficial femoral artery / popliteal artery.The 5.5 x 120 mm armada 18 dilatation catheter was prepared for use and no issues were noted.The device was advanced into the patient anatomy.The dilatation catheter balloon was inflated three times.After the third inflation to 14 atmospheres, the balloon ruptured.The rupture was large and the balloon separated from the device.The device was removed without difficulty and all separated pieces were removed without additional intervention.There was no adverse patient effect or clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual analysis was performed on the returned device.The reported material rupture and material separation were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints reported from this lot.The investigation determined that the reported difficulties were likely due to circumstances of the procedure.Based on the reported information, and analysis of the returned unit, it is likely that the balloon rupture occurred as the result of interaction with the severely calcified lesion.Additionally, it is likely that the separation occurred as the ruptured balloon material was retracted into the sheath resulting in in the balloon material tearing radially and ultimately separating the distal portion of the balloon catheter as noted on the returned device.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|