| Model Number |
050-87222 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty pdx, integrated set shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the inside of the cassette door of their cycler after ending their pd treatment.The patient reported receiving an air detected in cassette alarm during fill 1 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient confirmed that the solution did get into the cycler.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was available.Upon follow up, the patient stated it is unknown where the fluid leak occurred.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient has received a new cycler which is working well and is continuing peritoneal dialysis therapy with no further issues.The cassette used by the patient is not available for return for physical evaluation by the manufacturer.The cycler was returned to the manufacturer for physical evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h6.
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the inside of the cassette door of their cycler after ending their pd treatment.The patient reported receiving an air detected in cassette alarm during fill 1 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient confirmed that the solution did get into the cycler.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was available.Upon follow up, the patient stated it is unknown where the fluid leak occurred.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient has received a new cycler which is working well and is continuing peritoneal dialysis therapy with no further issues.The cassette used by the patient is not available for return for physical evaluation by the manufacturer.The cycler was returned to the manufacturer for physical evaluation.
|
| |
|
Search Alerts/Recalls
|