Model Number 71420170
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Laceration(s) (1946); Pain (1994); Loss of Range of Motion (2032); Osteolysis (2377); Swelling/ Edema (4577)
Date of Event 11/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tkjr surgery had been performed on (b)(6) 2012, the patient reported an unknown problem.The adverse event was addressed with a revision surgery on (b)(6) 2021 to replace the tibial baseplate, femoral component and the insert.Additionally was observed that there was no cement on the back of the implant, indicating that the product had failed had the cement/implant interface.The condition of the patient is unknown.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a tkjr surgery had been performed on (b)(6) 2012, the patient reported suffering from major knee effusion, lysis, poor rom and pain.The adverse event was addressed with a revision surgery on (b)(6) 2021 to replace the tibial baseplate, femoral component and the insert.Additionally was observed that there was no cement on the back of the implant, indicating that the product had failed had the cement/implant interface.The patient is on early stage of recovery.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the reported ¿major knee effusion, lysis, poor rom and pain¿ may be consistent with the reported failure of the cement/implant interface; however, with the limited information provided the clinical root cause of the reported clinical reactions/ events cannot be confirmed.It cannot be concluded that the reported events/clinical reactions were associated with a mal performance of the implant or implant failure.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes could be alignment, fit/size of device used, failure of the cement/implant interface, surgical complications, traumatic injury or wear.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II CMT TIB SIZE 6 LEFT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key13022487
Report Number1020279-2021-08709
Device Sequence Number14696675
Product Code JWH
UDI-Device Identifier03596010206282
UDI-Public03596010206282
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2022
Device Model Number71420170
Device Catalogue Number71420170
Device Lot Number12DT17924
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2021
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date12/16/2021
Supplement Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/23/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-