| Model Number |
OTV-S7H-1NA-10E |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility reported to olympus that the image was flickering.There was no patient injury, associated with the problem, reported to olympus.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned to olympus and evaluated.The reported problem was confirmed and the cause isolated to a faulty cable unit.The investigation is ongoing and the root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation, the investigation confirmed the image did not display due to a faulty cable unit.The specific root cause of the faulty cable unit could not be determined at this time.Olympus will continue to monitor field performance for this device.
|
| |
|