(b)(6) 2021, confirmed that the image was not displayed when inspecting the equipment, and there was a little noise.The tip generates heat from the curved tip.No major trauma or leaks.This event occurred at the time of before delivery.There was no report of patient harm.
|
Evaluation summary: we checked the returned unit and confirmed the ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the impurities is attached to the bending rubber , and the impurities or filth is attached to the insertion flexible tube (ift); however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.Correction information: type of report: follow up #1.Additional information: device manufacture date.
|