H10: manufacturing review: a complaint history review was performed.This is the second complaint reported for this lot number.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation.The returned sheath exhibited the presence of white fibers on the sheath tip and also on a 15mm section of the sheath located 45mm from the distal tip.Hydrophilic coating absent on sections of the sheath.The white fibers contamination was likely transferred to the sheath during device preparation by the hcp.The white fibers contamination is unlikely to be manufacture related as 100% visual inspection for device defects is performed during production.The result of the investigation is confirmed for the reported contamination issue.The definitive root cause for the reported contamination issue could not be determined based upon the available information received from the field communications and device evaluation.Labeling review: instructions for use for the halo one thin-walled guiding sheath was reviewed and the following sections are applicable: warnings: 1.Contents supplied sterile using ethylene oxide (eto).Non-pyrogenic.Do not re-use, reprocess or re-sterilize.This device is intended for single use only.2.Do not resterilize.After resterilization, the sterility of the product is not guaranteed because of an indeterminable degree of potential pyrogenic or microbial contamination which may lead to infectious complications.Cleaning, reprocessing and/or resterilization of the present medical device increases the probability that the device will malfunction due to potential adverse effects on components that are influenced by thermal and/mechanical changes.3.This device has been designed for single use only.Reusing this medical device bears the risk of cross-patient contamination as medical devices ¿ particularly those with long and small lumina, joints, and/or crevices between components ¿ are difficult or impossible to clean once body fluids or tissues with potential pyrogenic or microbial contamination have had contact with the medical device for an indeterminable period of time.The residue of biological material can promote the contamination of the device with pyrogens or microorganisms which may lead to infectious complications.4.Visually inspect the packaging to verify that the sterile barrier is intact.Do not use if the sterile barrier is opened or damaged.Precautions: 1.The halo one thin-walled guiding sheath shall only be used by trained physicians.Access procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment and / or ultrasound.6.Carefully inspect the sheath prior to use to verify that the sheath has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.21.In order to activate the hydrophilic coating, it is recommended to wet the halo one thin-walled guiding sheath with heparinized saline solution immediately prior to its insertion in the body.Failure to activate the coating may lead to sub-optimal trackability of the sheath.To maintain lubricity this surface must be kept completely wet.Directions for use: 5.In order to activate the hydrophilic coating, where labeled, it is recommended to wet the halo one thin-walled guiding sheath with sterile saline solution immediately prior to its insertion in the body.H10: d4(expiry date: 10/2022), g3.H11: h6(method, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|