LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-50-00
Medical Device Problem Code Failure to Auto Stop (2938)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure is intermittent.A video of the failure was provided and analysis revealed that despite rotating the knob to zero rpm, the indication of the rpms displayed on roller pump is not zero.The serial read-out of the pump has been taken and will be analyzed.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova (b)(4) received a report that a more than one pump used on a s5 system have a tendency to pump creep.Reportedly, the pumps correctly stop when the speed control knob is set to sero rpm but they start to turn again spontaneously.If the pump is stopped through the push button on the touchscreen, the reported issue doesn't occur.There was no report of patient injury.
 
Additional Manufacturer Narrative
H.10: read-out analysis did not reveal deviations.No hardware / software issues have been stored during the day of the fse visit on site, during which the issue could be reproduced.Currently the customer is likely using the pumps stopping them by pressing the stop function button on the touchscreen.Investigation still on going.
 
Event or Problem Description
See initial report.
 
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
H.10: affected roller pumps have been removed from service and loaners device have been provided to the customer.Through follow-up communication, livanova learned that the reported issue could be reproduced in approximately 1 out of 10 attempts.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
H10: the most likely root cause of the reported event is to be traced back to a malfunction of the shaft angle encoder.In detail, to the variability of the involved components with respect to the clinical use (the issue is reproducible only when rapidly turning the speed knob to zero).The shaft angle encoders were replaced by the technical department before shipping the roller pumps back to the customer.
 
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
H.10: two roller pumps affected (serial numbers: (b)(6) were returned back to livanova for product analysis.Investigation confirmed the reported issue.When turning rapidly the speed control knob, it can occur that, despite the bip sound, the rpm displayed is 1 (and not zero) and the pump runs accordingly.When the knob is turned very slowly and rpm are set to 0, the bip sound is not present and the rpm rapidly became 1 (and the pump runs accordingly).There are no errors stored in the pumps read-out.The same behavior was reproduced on both the roller pumps.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key13023921
Report Number9611109-2021-00710
Device Sequence Number17355619
Product Code DTQ
UDI-Device Identifier04033817900900
UDI-Public010403381790090011200907
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48-50-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2021
Supplement Date Received by Manufacturer01/19/2022
03/29/2022
10/19/2022
11/30/2022
Initial Report FDA Received Date12/17/2021
Supplement Report FDA Received Date02/16/2022
04/26/2022
11/17/2022
12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-