COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGTRS45AMT
Medical Device Problem Codes Misfire (2532); Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during an intestinal resection in an open abdominal surgery, the jaws did not open because the sheet of reinforcement material did not disengage from the jaws after squeezing the handle and performing the firing.Suture was cut with scissors to resolve the issue.No patient injury.Medtronic's initial evaluation of the incident is that reload was partially fired and the investigation detected an the buttress was torn and dislodged from the distal end on the anvil side.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the returned device to be pre-fired and engaged in interlock.Functional testing required the interlock to be overridden and the reload was applied to test media.The reload interlock was tested and found to function properly.It was reported that the reinforcement material did not release from the device as expected after firing.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13024242
Report Number1219930-2021-05248
Device Sequence Number14812919
Product Code GDW
UDI-Device Identifier10884521543645
UDI-Public10884521543645
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberSIGTRS45AMT
Device Catalogue NumberSIGTRS45AMT
Device Lot NumberN1F0926Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2021
Initial Report FDA Received Date12/17/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-