H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed; however, the root cause was not identified.The product returned for evaluation was one photograph which depicted a tls 3cg stylet.The visible region of the device included the polyimide region.Two sharp kinks were evident between the transition and distal tip.While stylet damage was obvious in the submitted photograph, inspection of the photograph was insufficient to identify the cause of the damage.Consequently this complaint is confirmed as ¿cause unknown¿ at this time.Potential contributing factors include stylet manipulation and insertion of the stylet past the catheter tip during catheter insertion.H3 other text : evaluation findings are in section h.11.
|